The plug-in module for analog input is designed to record voltages from -10 to +10 V and current signals from 0/4 to 20 mA. Laboratory devices such as thermostats, stirrers or measuring sensors can be connected to the four potential-free and EMC-resistant inputs. Sensors with two- and four-wire connections can be connected directly to a module socket, i.e. without an external power supply. The connection only takes a few seconds as there is no need to connect a power supply.
The module is particularly suitable for recording analog signals such as
Signal | 0/4...20 mA -10...+10 V |
Connection type | 6-pin socket, DIN 45322 |
Resolution (effective) | 20 Bit |
Measuring range | Voltage: -0.1...+0.1 V, -1...+1 V, -10...+10 V Current: 0/4...20 mA |
Absolute measurement uncertainty from final value (within the maintenance interval) | -10...+10 V: 0,05 % -10...+10 V: 0,05 % -0.1 V...+0.1 V: 0.05 % 0/4...20 mA: 0.15% |